Place the SiC/1.5 calibration sample on the flat horizontal working area under the AFM probe.
Approach the AFM probe to the sample surface and make topography scanning in the height measure mode using the scan size of about 10 um (Fig. 1).
· After obtaining good quality AFM-image of the sample surface with several steps use the software filter to flatten image so that every single step becomes horizontal (Fig. 2). Choose the area on AFM-image for obtaining height spectra by using possibilities of AFM software. Please, choose the area with maximum number of steps for better statistics. After obtaining height spectra with peaks corresponding to each step, measure the interpeak distances. Note that distances between neighboring peaks may vary a little (see Fig.3), so it is useful to average distances between peaks by measuring distance between far standing peaks and dividing the measured value by the number of included interpeak distances (A-A on Fig. 3). Change the scanner calibrating constant while average interpeak distance becomes 1.5 nm.